Cone beam computed tomography for dental and maxillofacial imaging: technique improvement and low-dose protocols

Beatrice Feragalli¹, Osvaldo Rampado¹, Cecilia Abate², Monica Macrì², Felice Festa²

  1. Department of Radiology, University of Foggia, Viale Luigi Pinto 1, 71100, Foggia, Italy
  2. Department of Radiology, Scientific Institute Hospital “Casa Sollievo Della Sofferenza”, Viale Cappuccini 1, 71013, San Giovanni Rotondo, FG, Italy

Abstract

Background

The aim of this study was to evaluate images quality and radiation doses of Cone Beam Computed Tomography (CBCT) for dental and maxillofacial imaging testing five different acquisition protocols.

Materials and Method

Dose measurements of different acquisition protocols were calculated for Pax Zenith three-dimensional (3D) Cone Beam (Vatech, Korea) and for conventional orthopantomography (OPT) and cephalometric skull imaging Ortophos (Sirona Dental Systems, Bernsheim, Germany). The absorbed organ doses were measured using an anthropomorphic phantom loaded with thermoluminescent dosimeters at 58 sites related to sensitive organs. Five different CBCT protocols were evaluated for image quality and radiation doses. They differed in FOV, image resolution, kVp, mA, acquisition time in seconds and radiation dose. Measurements were then carried out with the orthopantomograph. Equivalent and effective doses were calculated.

Results

The reference protocol with large FOV, high resolution quality images, 95 kVp, 5 mA and acquisition time of 24 s resulted in a DAP value of 1556 mGy cm2 instead the protocol with reduced kVp from 95 to 80 kVp translated into a value of DAP inferior to 35% (from 1556 to 1013 mGy cm2). Going from a high resolution to a normal resolution, there was a reduction of the acquisition time to 15 s which allowed further dose reduction of approximately 40% (628 mGy cm2); this protocol resulted in a value of effective dose of 35 microSievert (μSv). Moreover, the effect of changing FOV has been evaluated, considering two scans with a reduced FOV (160 × 140  and 120 × 90 mm, respectively).

Conclusion

CBCT low-dose protocol with large FOV, normal resolution quality images, 80 kVp, 5 mA and acquisition time of 15 s resulted in a value of effective dose of 35 microSievert (μSv). This protocol allows the study of maxillofacial region with high quality of images and a very low radiation dose and, therefore, could be proposed in selected case where a complete assessment of dental and maxillofacial region is useful for treatment planning.

Keywords: CBCT, maxillofacial, radiation dose